Instrument types

Rheometer is a mechanical spectrometer that is capable of subjecting a sample to either a dynamic (sinusoidal) or steady (linear) shear strain (deformation). Then measuring the resultant torque expended by the sample in response to the shear strain (shear strain is applied by the motor- torque is measured by the transducer). In the dynamic mode, the motor begins all tests at the motor zero position, and drives symmetrically about motor zero to the strain commanded by the software. The motor is labeled with graduations indicating 0.1, 0.25, and 0.5 radians from motor zero. The maximum angular deflection of the motor is 0.5 radians from either side of motor zero. In the steady mode, the motor can begin a test from any position, rotating either clockwise or counterclockwise as specified by the user using software. Available geometry:

 


Geometry Application Advantage Disadvantage
Cone/plate Fluids, melts Viscosity > 10 mPas True viscosity Temperature ramp difficult
Parallel Plate Fluids, melts Viscosity > 10 mPas Easy handling, temperature ramp Shear gradient across sample
Couette Low Viscosity samples <10 mPas High shear rate Large sample volume

 

 

Plates size: Low viscosity (milk)- 60 mm geometry, medium viscosity (honey)- 40 mm geometry, high viscosity (caramel)- 20 mm geometry.

Equipment

Instrument: LC-ARES Test Station from Rheometric Scientific.

Temperature: oven 22oC - 200oC

.